Zobrazit stránku 14
Watlow MicroDIN 2.5
Communications
Setup
MicroDIN, A Modbus
Remote Terminal Unit (RTU)
RTU
The MicroDIN uses Modbus
RTU (remote terminal unit) protocol to read and
write to registers that can be viewed or changed from a personal computer.
Sending ASCII text commands to the MicroDIN will not work.
Access Via Registers
Each MicroDIN parameter has a corresponding Modbus
register and access
privileges. The value of each parameter is stored in a Modbus
register. The
access privilege — read only, write only or read/write — determines whether you
can change and/or view a parameter’s value. Chapter 5 explains some of the
interactions between parameters. The tables at the end of this chapter show the
parameters and their setup order.
MicroDIN Supports...
Multiple Register Reads
MicroDIN supports Modbus
Commands 03 and 04 read of a 1- to 32-register
block. Responses from -1999 to 9999 are valid data for up to 32 register writes.
Modbus
32000 Errors
A Modbus
read command response of -32000 indicates that a register is not
implemented; -32001, register not active; or -32002, not read accessible.
Single Register Writes
MicroDIN supports a Modbus
write command (06) to a single register.
Multiple Register Writes
MicroDIN supports a Modbus
write command (16) for multiple writes as long
as the data length is 1.
Writing to address 0 broadcasts to all devices on the network.
Write Exceptions
MicroDIN supports write exceptions: 01, illegal function; 02, illegal register; 03,
illegal data.
Loop Back
MicroDIN supports a Modbus
write command 08, “loop back” to echo a
message.
Compatible Software
You will need compatible software to interact with a MicroDIN controller. One
such program is included with the MicroDIN Startup Kit. If you are accustomed
to using a Modbus
, see the Appendix for information to help write your own
Modbus
software application.
çCAUTION:
Avoid writing
continuously to
EEPROM memory.
Continuous writes
may result in
premature control
failure, system
downtime and
damage to
processes and
equipment.
See Disable Non-
volatile Memory
(System Group),
Ch. 5.