0600-0015-0000 Rev L $15.00
January 2002
Supersedes: 0600-0015-0000 Rev K
Users Manual
Registered Company
Winona, Minnesota USA
TOTAL
3 Year Warranty
CUSTOMER
SATISFACTION
Series 935A
Temperature Controller
with Countdown Timer
1241 Bundy Boulevard, P.O. Box 5580, Winona, Minnesota USA 55987-5580
Phone: +1 (507) 454-5300, Fax: +1 (507) 452-4507, http://www.watlow.com
U.S. English
Zobrazit stránku 0
1 2 3 4 5 6 ... 67 68

Shrnutí obsahu

Strany 1 - Series 935A

0600-0015-0000 Rev L $15.00January 2002Supersedes: 0600-0015-0000 Rev KUser’s ManualRegistered Company Winona, Minnesota USA TOTAL3 Year WarrantyCUST

Strany 2 - Your Comments

8 Watlow Series 935A User’s ManualRespond to a simple errorDisplay Probable Cause Recommended ActionReversed thermo- Change the sensorcouple connectio

Strany 3 - Where to find it…

Watlow Series 935A User’s Manual 9Software Organization• The Series 935A has three primary menus in addition to a normal display.• The software revert

Strany 4 - Introduction

10 Watlow Series 935A User’s ManualThe Series 935A Operations Menu is the first menu you encounter when you press the . and , keys simultaneously for

Strany 5 - Dual Output

Watlow Series 935A User’s Manual 11Table 4 - Operations Menu Overview To enter the Operations Menu, press the . and , keys simultaneously for three se

Strany 6 - Read or change

12 Watlow Series 935A User’s ManualChoose the PID StrategyYou may rely solely on the Auto-tune [`Aut] function (p. 32) and factory defaults to determi

Strany 7

Watlow Series 935A User’s Manual 13Set the PID Menu Values• Proportional Band, Heat and Cool [Pb`h] and [Pb`c]: Select a value (degrees) to set up ban

Strany 8 - Begin Controlling

14 Watlow Series 935A User’s ManualTable 5 - PID Menu Overview Set-Up HeatProportional Band Heat - Select a heat proportional band value.[```0] [`999]

Strany 9 - Change Set Point

Watlow Series 935A User’s Manual 15Set-Up GeneralIntegral Function - Select an integral value.[`)00] [9(99]0.00 to 99.99 minutes/repeatDerivative Func

Strany 10 - Respond to a simple error

16 Watlow Series 935A User’s ManualThe Configuration Menu is the second sub-menu in the Operations Menu. Use it to set Inputs, Ranges, Output Types, A

Strany 11 - Learn the Software Map

Watlow Series 935A User’s Manual 17NOTE:Access to Configuration Menu varies with lockout function. See p. 19.Alarm Hysteresis - Choose alarm switching

Strany 12 - Learn the Operations Menu

Safety Information in this ManualNote, caution and warning symbols appear throughout this book to draw your attention toimportant operational and safe

Strany 13

18 Watlow Series 935A User’s ManualThe Series 935A offers three different security, or “lockout,” options. Set up one or all three lockout op-tions in

Strany 14 - Learn the PID Menu

Watlow Series 935A User’s Manual 19View Process Yes Yes Yes Yes Yes Yes YesView Set Point Yes Yes Yes Yes Yes Yes YesChange Set Point No Yes No Yes Ye

Strany 15 - Set the PID Menu Values

20 Watlow Series 935A User’s ManualSetting Inputs and OutputsKey Input/Output Set-up Information• All initial input and output set-up occurs in the Co

Strany 16 - Learn the PID Menu - Details

Watlow Series 935A User’s Manual 21Input Range InformationJ t/c: 32 to 1382°F or 0 to 750°CK t/c: -328 to 2282°F or -200 to 1250°CT t/c: -328 to 662°F

Strany 17 - Set-Up General

22 Watlow Series 935A User’s ManualTo Set Up Remote Set Point…1. Wire the control per the example below and the information on p. 42-45.2. Go to [CnfG

Strany 18 - Learn the Configuration Menu

Watlow Series 935A User’s Manual 23To Set Up Front Panel Lock…1. Install an external switch.2. Wire the control per the example below and the informat

Strany 19

24 Watlow Series 935A User’s ManualConfiguration Menu[``In]Input Type[`C_F]Celsius/Fahrenheit[``rL]Input Range Low[``rh]Input Range High[`Ot1]Output 1

Strany 20 - Using Lockout Functions

Watlow Series 935A User’s Manual 25Table 10 - Alarm FunctionsSet PointDeviationnormallyclosedDeviationnormallyopenProcessnormallyclosedProcessnormally

Strany 21 - [`Ot1] [SLOC] [`tAG]

26 Watlow Series 935A User’s Manual1. Plan an alarm strategy. What do you want to happen when an alarm occurs?2. Wire the appropriate control output,

Strany 22 - Setting Inputs and Outputs

Table 11 (below) and Table 10 (p. 25) illustrate the Series 935A alarm features.Table 11 - Alarm FunctionsSETLatching alarmSilence alarmWith this sele

Strany 23

Watlow Series 935A User’s Manual 1Where to find it…Table of ContentsItem ... PageIntroductio

Strany 24 - Set Up Remote Set Point Input

28 Watlow Series 935A User’s ManualConfiguration Menu[``In]Input Type [`C_F]Celsius/Fahrenheit[``rL]Input Range Low[``rH]Input Range High[`Ot1]Output

Strany 25 - Set Up Front Panel Lockout

Watlow Series 935A User’s Manual 29Table 12 - Series 935A Timer Functions/SettingsDelay ONDelay OFF Signal ON Signal OFFOutput 2Timer OutputFunction;C

Strany 26 - Learning Alarms

30 Watlow Series 935A User’s Manual3015 4501. Plan a timer strategy.2. Wire the Output 2 control output, associated switching devices and annunciators

Strany 27 - [dEnc] [dEno] [Prnc] [Prno]

Watlow Series 935A User’s Manual 313015 450Convection Oven ApplicationScenarioA master chef bakes bread at 350°F for 30 minutes. He wants the oven at

Strany 28 - Setting Alarms

Auto-tuningAuto-tune automatically sets PID parameters for your system.1. Press . and , for three seconds.2. You'll see [`Aut].3. Press and hold

Strany 29 - Table 11 - Alarm Functions

Watlow Series 935A User’s Manual 33Fine Tune the PID Settings1. Set [Pb`h] and [Ct`h] in degrees.2. If Proportional Band Heat [Pb`h] = 0, Set Hysteres

Strany 30 - Learn the Countdown Timer

34 Watlow Series 935A User’s ManualCalibrating the 935ACalibration requires a precision millivolt source with thermocouple compensation, an adjustable

Strany 31

Watlow Series 935A User’s Manual 35Thermocouple and Remote Set Point Input Field Calibration ProcedureEquipment Required:• Type “J” Reference Compensa

Strany 32 - Setting the Countdown Timer

36 Watlow Series 935A User’s ManualErrors and TroubleshootingSet up an input failure operation mode at the [FAIL] prompt in the [CnFG] menu; choose bu

Strany 33 - Timer Example

Watlow Series 935A User’s Manual 37To Troubleshoot Sensor• Remove sensor wires from Terminals 1 and 2.• For a thermocouple sensor Series 935A, place a

Strany 34 - Set Point

2 Watlow Series 935A User’s ManualWelcome to the Watlow Series 935A!General DescriptionThe Series 935A is a temperature controller with countdown time

Strany 35 - Fine Tune the PID Settings

38 Watlow Series 935A User’s ManualMounting22.2mm(0.874 in.)+0.3mm,-0.0mm(+0.012 in.,-0.000 in.)12.7mm(0.500 in.)12.7mm(0.500 in.)57.7mm(2.272 in.)34.

Strany 36 - Calibrating the 935A

Watlow Series 935A User’s Manual 39DimensionsFigure 11- Series 935A Dimensions53.1mm (2.09 in.)29.0mm(1.14 in.)Panel Cutout 44.96mm 101.6mm (4.00 in.

Strany 37

40 Watlow Series 935A User’s ManualInstallation Procedure1. Make a panel cutout using the dimensions in Figure 9, p.38.2. Insert the 935A into the cut

Strany 38 - [FAIL} [bPLS]

Figure 14- Terminal Block Removal Procedure1. Press in on sides of cover to release the terminal cover hooks.2. Move your grip rearward slightly, then

Strany 39

42 Watlow Series 935A User’s ManualWiring a 935AFigure 15 - Wiring the Series 935A∫WARNING: All electrical wiring and fusing must conform to local and

Strany 40 - Mounting

Watlow Series 935A User’s Manual 43∫WARNING: All electrical wiring and fusing must conform to local and national electric codes. Contact local authori

Strany 41 - Dimensions

44 Watlow Series 935A User’s ManualSystem Wiring ExamplesFigure 16 - Series 935A System Wiring Examples∫WARNING: All electrical wiring and fusing mus

Strany 42 - Installation Procedure

Watlow Series 935A User’s Manual 45Figure 17 - Series 935A Ladder Diagram Wiring Example∫WARNING: All electrical wiring and fusing must conform to lo

Strany 43 - Terminal Block Removal

46 Watlow Series 935A User’s ManualGlossaryAlarmA condition, generated by the controller, indicating that the pro-cess has exceeded or fallen below th

Strany 44 - Wiring a 935A

Watlow Series 935A User’s Manual 47Ready AcknowledgeA Series 935A countdown timer start choice that pre-initiates thetimer with a down key press, and

Strany 45 - Internal Circuitry

Watlow Series 935A User’s Manual 3Output 1Output 2RDY21SETSingleInputDual OutputSingle InputType J, K, T, N, S, E Thermocouple, 1° RTD, or 0.1° RTDOut

Strany 46 - System Wiring Examples

48 Watlow Series 935A User’s ManualIndexAAuto-tune [`Aut] 10, 32AlarmsAlarm High [`AhI] 10, 24Alarm Low [`ALO] 10, 24Alarm Type [AltY] 24Clearing Alar

Strany 47

Watlow Series 935A User’s Manual 49SSet Point Change 7Local [`L-r] 10Lock [SLOC] 18Range High [``rh] 20Range Low [``rL] 20Remote Set Point [`rSP] 22Re

Strany 48 - Glossary

50 Watlow Series 935A User’s ManualSpecificationsSpecifications (2191)Control Mode• Auto-tune PID.• PID, PD, PI, on/off.• Countdown timer; hours:minut

Strany 49

Watlow Series 935A User’s Manual 51Ordering a 935AOrdering Information (2192)Single thermocouple or RTD (DIN) input, dual output, singlie display temp

Strany 50 - [`CAL] 13

52 Watlow Series 935A User’s ManualDeclaration of ConformitySeries 935WATLOW Winona, Inc.1241 Bundy BoulevardWinona, Minnesota 55987 USADeclares that

Strany 51

Watlow Series 935A User’s Manual 53Notes

Strany 52 - Specifications

54 Watlow Series 935A User’s ManualNotes

Strany 53 - Ordering a 935A

Watlow Series 935A User’s Manual 55935A Software MapOperations MenuPrompt Range Default Hidden if * Your Settings*NOTE:All prompts, except[``SP] and [

Strany 54 - Series 935

56 Watlow Series 935A User’s ManualOperations MenuPrompt Range Default Hidden if * Your Settings[`tA9] = [P```];or if [Pb`h] ≠ 03°F or 2°C[```1] [`999

Strany 55

Watlow Series 935A User’s Manual 57PID MenuPrompt Range Default Hidden if * Your Settings[`tA9] = [P```];or if [Pb`c] ≠ 03°F or 2°C[```1] [`999][```1]

Strany 56

4 Watlow Series 935A User’s ManualRead or changeYou can simply:• Read the normally displayed actual temperature,or…• Press and hold ß to read the set

Strany 57 - 935A Software Map

58 Watlow Series 935A User’s ManualPID MenuPrompt Range Default Hidden if * Your Settings[`tA9] = [P```]0°F or 0°C[-999] [9999][-9(9] [99(9]Select a v

Strany 58

Watlow Series 935A User’s Manual 59Configuration MenuPrompt Range Default Hidden if * Your Settings[`tA9] = [``C`]heat[hEAt] [COOL][ALM] [`rSP][`FPL]

Strany 59

60 Watlow Series 935A User’s ManualConfiguration MenuPrompt Range Default Hidden if * Your Settings[`tA9] = [``C`];or if[`Ot1] and/or [`Ot2] ≠ [ALM]3°

Strany 60 - [`tA9] = [P```];

Watlow Series 935A User’s Manual 61Configuration MenuPrompt Range Default Hidden if * Your Settings[`tA9] = [``C`];or if[`Ot2] ≠ [thM] or[tMS]0°F or 0

Strany 61

62 Watlow Series 935A User’s ManualConfiguration MenuPrompt Range Default Hidden if * Your SettingsAlways appearsP[PCOA] [`COA][PCO`] [`CO`][PC`A] [`C

Strany 62

Watlow Series 935A User’s Manual 63Notes

Strany 63

64 Watlow Series 935A User’s ManualNotes

Strany 64

About Watlow WinonaWatlow Winona is a U.S. division of Watlow Electric Manufacturing Company, St. Louis,Missouri, a manufacturer of industrial electri

Strany 65

How to Reach UsQuality andMission Statement:Watlow Winona will be theworld’s best supplier ofsuperior measurement andcontrol products, services,and sy

Strany 66

RDY21SETSeven-segmentalphanumeric display:• Shows process value,set point information, time, or• Shows prompt name or value, depending on the key comb

Strany 67 - Warranty

6 Watlow Series 935A User’s ManualBegin Controlling1. Apply power to the system.A properly-wired Series 935A will begin controlling the thermal system

Strany 68 - How to Reach Us

Watlow Series 935A User’s Manual 7Change Set PointRDY21SETYour Series 935A displays the actual process temperature when it comes from the factory. You

Komentáře k této Příručce